The fatigue life of the unit is 10E9 full stress reversed cycles at the capacity of the cell. Overload capability is 300% of the capacity before mechanical failure of the cell
Side load resistance is 40% of the capacity
Dynacell is a unique load measurement device which automatically compensates for load errors induced through inertia of the load string components. A method of dynamic inertia compensation is essential for all forms of dynamic testing.
The 2527 Series load cells are designed for use with Dynamic Testing Systems; offering exceptional performance with the ability to measure forces as low as 1/250th of the force capacity to an accuracy of 0.5% of reading. Automatic transducer recognition and electrical calibration, makes them easy to use. The load cells can withstand loads up to 150% of their force capacity without damage and 300% without mechanical failure. All Instron load cells are individually temperature-compensated and tested for accuracy and repeatability on calibration apparatus that is traceable to international standards, with a measurement uncertainty that does not exceed one-third of the permissible error of the load cell.