Наш сайт использует куки-файлы для того чтобы обеспечить максимальные удобства пользователям. Продолжая просматривать наш сайт Вы соглашаетесь с использованием куки-файлов. Подробнее
2525-815 | 2525 Series Drop-Through Static Load Cell (2.5 N)
Static Load Cell: 2.5 N (250 g, 0.5 lb), Type OOf, Of
Tension/Compression
Type OO female fitting (6 mm connection with 2.5 mm diameter clevis pin)
Includes adapter to Type O female fitting (12 mm connection with 6 mm diameter clevis pin)
Drop through, bolt on design with self identification
and electrical calibration
Linearity: +/- 0.25% of reading over a range of 0.4%
to 100% of capacity
Repeatability: +/- 0.25% of reading over a range of
0.4% to 100% of capacity
Hysteresis: 0.2% of maximum reading in cycle
Creep: 0.2% of reading over 3 minutes
Output Sensitivity: 2.0 millivolts per volt +/- 20%
at full rated capacity
Bridge Resistance: 350 ohms +/- 5%
Overload Protection: 150% of capacity without
permanent zero shift. 300% of capacity or 5 kN
(500 kg, 1000 lb) whichever is greater without
mechanical failure
Zero Stability: 0.001% of full rated output per hour
(isothermal conditions)
Temperature Compensation: 0 to 50 degrees
Celsius (32 to 122 degrees Fahrenheit)
Temperature Effect on Zero and Span: 0.002% of
capacity per degree Celsius